Labratories

Tokyo Laboratory

EMC for wide range of electronics devices -- Emission and Immunity Testing

Accreditations:VLAC, BSMI, VCA, Mazda

Tokyo Lab. is preferably close to Tokyo metropolitan area, taking only 30-40 minutes by train, and is also easily accessible by car only to require 15-30 minute drive from each exit of Tokyo's two main express ways, Tomei and Chuo.

Equipped with the semi-anechoic chambers for testing at 3m/10m measurement distance, our test engineers are all well trained and experienced to offer you high-quality EMC testing services for all the electronic devices including medical / measurement / IT devices as well as automotive components.

Six customers can use the test facilities at the same time with separate entrances to carry in their products, which keeps customers' top confidential matters secure and safe.

inside the 10m chamber
Semi-anechoic chamber for testing at 10m measurement distance.
entrance: 10m chamber
Wide entrance
of the semi-anechoic chamber.
elevator at B1: 10m chamber
Elevator installed under the chamber.
3m chamber
Semi-anechoic chamber
for testing at 3 m measurement distance.
shielded room
Shieled rooms

Testing Facilities

Anechoic chamber
for testing at 10 m measurement distance
Registered to FCC and VCCI / CE marking possible 1
Anechoic chamber
for testing at 3m
measurement distance

Registered to FCC and VCCI / CE marking possible
EMC testing for Automotive, Military, Aerospace electronics

1
Small semi-anechoic chamber EMC testing for Automotive, Military, Aerospace electronics 1基
シールドルーム EMC testing for Automotive, Military, Aerospace electronics(Conducted)
3基
200V/mの電界強度試験10kHz-18GHz
Testing for 200V/m electric field strength at 10KHz-18GHz
Complete set of the test equipment suitable for VCCI / FCC / CE / MIL / JASO and others
Various power supplies for EUT: AC40~400Hz/9kVA/3φ、DC60V/60A, etc.

Site Specifications

Anechoic Chambers(3)

Room No.  No.1  No.5 No.6
Specs  10m Anechoic
Chamber 
3m Anechoic
Chamber
Small Anechoic
Chamber 
Facilitiy Dimensions
(W×L×H)(m)
14(12)×21×9.5 6×9×5.5 5x6x3
Product Entrance Dimensions
 (of anechoic chambers) (W×H)(m)
1.8×1.8 1.2×2 1×2
Turn-table Diameter
(m)
3.1m 1.5m -
Load Capacity (t) 1 0.5 0.5
Power Supply Single phase 6kVA
70~264V
50/60Hz
6kVA
70~264V
50/60Hz
6kVA
70~264V
50/60Hz
Three phase
Four lines
6kVA
70~264V
40~500Hz
6kVA
70~264V
40~500Hz
-
DC DC60V60A
DC60V60A
DC60V60A
Measurement Distance 1m/3m/10m 1m/3m 1m
EUT Dimensions that can be carried in (m) 2×1.8 1×1.8 1×1.8

Shielded Rooms(3)

No. No.2 No.3
No.4
Specs
Facilitiy Dimensions (W×L×H)(m) 4×5×3 5×6×3 5×6×3
 Product Entrance Dimensions
 (of anechoic chambers)
(W×H)(m)
1×2 1×2 1×2
Load Capacity (t) 0.5 0.5 0.5
Power Supply Single phase 6kVA
70~264V
50/60Hz
6kVA
70~264V
50/60Hz
6kVA
70~264V
50/60Hz
Three phase Four lines 6kVA
70~264V
50/60Hz
6kVA
70~264V
50/60Hz
6kVA
70~264V
50/60Hz
DC DC60V30A DC60V30A DC60V30A
EUT Dimensions that can be carried in (W×L×H)(m) 1×1.8 1×1.8 1×1.8

EMC Test specifications

Emission Measurement

Test Item Specifications
Electric Field Strength Measurement 10kHz~26GHz(CISPR Pub.25 Level5対応)
Magnetic Field Strength Measurement DC~30MHz
Conducted Emission Voltage DC~108MHz
Mains Harmonic Measurement 100~240V/16A
Voltage Variation, Flicker Measurement 100~240V/16A

Immunity Testing

Test Item Specifications
Electrostatic Discharge Test 200V~30kV
EN61000-4-2/ISO10605
Radiated Immunity Test Strip Line & Antenna method
  • Frequency Range:10kHz~18GHz
    E-field strength:200V/m (non-modulation)
TEM Cell Frequency Range:10kHz~400MHz
Strip Line Frequency Range:10kHz~200MHz
Proximity Antenna Each frequency between 27MHz~1950MHz
EFT/Burst Immunity Test Maximum Test Voltage: ±4.4kV
Lightning Surge Immunity Test Maximum Test Voltage: ±6.6kV 1.2/50(8/20)μs
Conducted Immunity Test Maximum Test Voltage:30V 
Frequency Range:10kHz~2GHz
Induced Magnetic Field Immunity Test Maximum Test Voltage:30A/m 
Frequency Range:30Hz~100kHz
Power Voltage Variation Test DC~240V(Single phase) EN61000-4-11/JASO D001-94/, etc.
Low Frequency Conducted Immunity Test 5V(100W) Frequency Range:30Hz~250kHz
Conducted Transient Immunity Test ISO7637/JASO D001-94/, etc.
Damped Oscillatory Wave Immunity Test 1.5MHz(IEEE/ANSI C37.90a-1978)

* Please don't hesitate to ask anything you require.